Question

A push-pull capacitive displacement transducer is interfaced to a differential amplifier and an ADC as shown in the figure below.

Note that the bridge supply and the analog reference input for the ADC are derived from the same 10 V DC source.
(Gain of amplifier=40)

If the supply voltage to the bridge decreases by 5%, the sensitivity of the measurement system

A
decreases by 5%
No worries! We‘ve got your back. Try BYJU‘S free classes today!
B
does not change
Right on! Give the BNAT exam to get a 100% scholarship for BYJUS courses
C
increases by 5%
No worries! We‘ve got your back. Try BYJU‘S free classes today!
D
increases by 200%
No worries! We‘ve got your back. Try BYJU‘S free classes today!
Open in App
Solution

The correct option is B does not change
If the supply voltage to the bridge is reduced by 5% then the reference voltage to A/D converter also reduced by 5%.

So,

Vin=10V0.05×10=9.5V

Vref to A/D converter = 9.5 V

Then at full scale displacement
ΔCC=0.005

Bridge output voltage,
VOm=Vin2ΔCC

Push - Pull arrangement,
VOB=9.52×0.05=0.2375

Analog input to A/D converter,

(Vin)ADC=gain of amplifier×VOB

=40×0.2375=9.5V

So, at full scale input displacement analog input to A/D converter is equals to analog reference input, then digital display will be full scale. This means there is no change in sensitivity.