A 412 digital multimeter has the error specification of 0.5% of its reading + 10 counts. If a D.C. voltage of 50V is Read on its 200 V full scale. Then the maximum error that can be expected in the reading is ± _________ %.
Error due to 10 counts
=20020000×10=0.1V
Total error=0.10+0.25=0.35V