A 412 digital multimeter has the error specification of 0.5% of its reading + 10 counts. If a D.C. voltage of 50V is Read on its 200 V full scale. Then the maximum error that can be expected in the reading is ± _________ %.

A
7
No worries! We‘ve got your back. Try BYJU‘S free classes today!
B
0.7
Right on! Give the BNAT exam to get a 100% scholarship for BYJUS courses
C
70
No worries! We‘ve got your back. Try BYJU‘S free classes today!
D
0.07
No worries! We‘ve got your back. Try BYJU‘S free classes today!
Open in App
Solution

The correct option is B 0.7
Error due to 0.5% of reading
=0.5100×50=0.25V

Error due to 10 counts
=20020000×10=0.1V

Total error=0.10+0.25=0.35V

%error=0.3550×100=0.7%