A symmetrical square-wire voltage is applied to an average reading diode voltmeter having a scale calibrated in terms of rms value of a sine wave, the error in meter indication will be
A
1.11 %
No worries! We‘ve got your back. Try BYJU‘S free classes today!
B
11.1 %
Right on! Give the BNAT exam to get a 100% scholarship for BYJUS courses
C
21.1 %
No worries! We‘ve got your back. Try BYJU‘S free classes today!
D
0.45 %
No worries! We‘ve got your back. Try BYJU‘S free classes today!
Open in App
Solution
The correct option is B 11.1 % The deflection of the instrument depends upon the average value of current. The form factor of a sinusoidal wave is 1.11 and therefore average value of current fed to the meter for sinusoidal voltage supply is
Iav=Eav2R=Erms2×1.11×R=0.45ErmsR
The form factor of a square wave is 1 and with half wave rectification, the average value of current with square wave is