Question

A temperature - measuring instrument is modeled as a first order system with a time constant of 5s. The sensor of the instrument is placed inside an oil bath whose tempreture has a sinisoidal variation with amlitude of 10oC and a period of 20s around an average tempreture of 200 oC. The sinusoidal component at the output of the instrument will have amplitude.

A
0oC
No worries! We‘ve got your back. Try BYJU‘S free classes today!
B
5.37 oC
Right on! Give the BNAT exam to get a 100% scholarship for BYJUS courses
C
8.57 oC
No worries! We‘ve got your back. Try BYJU‘S free classes today!
D
10oC
No worries! We‘ve got your back. Try BYJU‘S free classes today!
Open in App
Solution

The correct option is B 5.37 oC
The frequently response of a first order system when it is subjected to the sinusoidal input is given as :

AoAi=1(1+ωτ)2

Given, τ= 5 sec

ω=2πT=2π20=π10

Ai=10oC

Ao=101+(π10×5)2=101+(3.142)2

=101+2.46=101.86=5.37o