An electromagnetic flow meter is used to measure the blood flow in a vessel. The diameter of vessel is 10 mm and the flux density across the electrode is π mm. If the output voltage for a particular flow rate is 1mV then the flow rate in mm3/s is ?


  1. 2.5

Open in App
Solution

The correct option is A 2.5
B=flux density,

l = diameter,

v =velocity,

V0 = output voltage

Q=v.A=v.π4l2

Q=V0Blπ4l2=V0πl4B

Putting the given values,

Q=1×103×π×10×1034π×103

Q=2.5×103

Q=2.5mm3/s